Parameters and specifications
- designed exclusively for measurements with LR8450/LR8450-01 current input units
- ambient temperature range in which measurements can be made with guaranteed specifications: -40°C to 85°C
| CT7812 | CT7822 | |
| frequency range | DC to 100 kHz | |
| max. input current | 3 A | 30 A |
| current ranges | 200 mA/2 A | 2 A/20 A |
| highest resolution | 0.1 mA | 1 mA |
| output | 100 mV/A | 10 mV/A |
| max. diameter of measuring wire | 5 mm | |
| cable length | 4 m | |
| interface | PL14 | |
| basic accuracy | DC± 0.3% rdg. ±0.1% f.s. | |
Downloads and links