Parameters and specifications
frequency range |
150 kHz to 30 MHz |
correction factor k1 (only valid for 22 mm cable diameter) | |
A3: |
20 dB |
frequency response correction factor k1 | |
A3: |
±1.2 dB |
power supply | |
internal |
6 x 1.2 V, NiMH battery, e.g. 1700 mAh |
current, type | 180 mA |
battery operation (typical) | 8 h |
cable size | |
max |
32 mm 5 mm |
connectors |
|
output to test receiver | BNC 50 Ω, female |
input for power supply | jack bush, ∅ 3.5 mm |
- capacitive voltage probe (CVP) is designed for measuring asymmetrical interference on cables based on the capacitive coupling principle
- provides the possibility to perform measurements without disconnecting the cable under test and thus prevents unwanted transients
- the main parameter of the CVP is the coupling factor, measured in a calibration system with an impedance of 50 Ω
- the capacitive voltage probe is specified in Section 5.2.2 and Annex G of CISPR 16-1-2.
- The capacitive voltage probe (CVP) is used for measurements at telecommunication ports for lines with more than four symmetrical pairs or for unsymmetrical lines