ModuLab XM ECS Electrochemical Test System

    • electrochemical test system
    • configurable platform for electrochemical applications such as measuring impedance in the μΩ range, e.g. for battery or fuel cells, or conversely for accurate characterization of high impedance corrosion layers
    • wide range of techniques includes CV, CC-CV charge/discharge, SW (Square-Wave) voltammetry, differential pulse voltammetry, linear sweep voltammetry and equivalent potentiometric techniques
    • accurate AC characterization through high oversampling of waveform generation for FRA (Frequency Response Analyses) includes measurements of impedance, admittance, permittivity/capacitance, electrical modulus, Mott-Schottky analysis, and more
    • test selection, measurement control and results display are managed by the intuitive PC-based XM-studio software
    • the XM ECS (Xtreme Measurement Electrochemical Test System) version of the ModuLab is designed for electrochemical research, but can also be extended for materials research or photovoltaic measurements
    • warranty: 2 years
    Solartron Analytical
    Other parameters
    Price on request
    Cat. no: ModuLabXMECS.w
    availability on request

    No item matches your selection.

    Parameters and specifications

    voltage range (compliance/polarization)

    ±8 V (PSTAT)
    ±30 V (HV30)
    ±100 V (HV100)
    ±20 V (HV30/HV100 + BOOSTER 2A)

    voltage resolution (RE) 1 μV (PSTAT)
    1.25 μV (HV30/HV100)
    current range

    ±300 mA (PSTAT)
    ±200 mA (HV30)
    ±100 mA (HV100)
    ±2 A (BOOSTER 2A)

    current resolution (WE)

    1.5 pA (PSTAT/HV30/HV100/BOOSTER 2A)
    0.15 fA (FEMTO AMMETER)

    sampling frequency

    1 MS/s (PSTAT)
    40 MS/s (FRA 300 kHz / FRA 1 MHz)

    maximum sample generation frequency

    64 MS/s (PSTAT)

    frequency analysis range 10 μHz to 300 kHz (FRA 300 kHz)
    10 μHz to 1 MHz (FRA 1 MHz)
    • two-, three- or four-wire connection modes
    • isolated (floating) measurement
    • large impedance measurement range
    • control modules:
      - XM PSTAT 1 MS/s - for time domain analysis (XM PSTAT AUX - choice of four auxiliary voltage channels)
      - XM FRA 1 MHz / FRA 300 kHz - for AC measurements (EIS)
    • slave modules for modification of signal parameters:
      - XM MHV30 / HV100 - high voltage options (30 V / 100 V)
      - XM FEMTO AMMETER - low current option
      - XM BOOSTER 2A - high current option (2 A)
    • multi-channel system configurable in combination with a material or PV research setup, whereby simultaneous measurements on the channels can each be controlled separately
    • modules are plug and play
    • SW XM-studio
      - allows the creation of automatic analysis sequences in time domain and AC techniques
      - offers a range of data analysis options such as curve fitting obtained from equivalent schemes
    • multisine FFT analysis speeds up the test performed over the entire frequency band
    • harmonic intermodulation analysis
    • availability of accessories (Corrosion/Tait/Flat Cell, High Power Boosters...)

    Downloads and links