Parameters and specifications
input module U7001 | input module U7005 | |
voltage* | max. 1000 VAC / 1500 VDC | max. 1000 VAC/DC |
sampling | 2.5 MHz | 15 MHz |
resolution | 16 bits | 18 bits |
max. number in analyzer | 8 | 8 |
measured frequency band | DC; 0.1 Hz to 1 MHz | DC; 0,1 Hz to 5 MHz |
*for voltage measurements up to 5000 V can be supplemented with an external VT1005 divider
- models with one to eight input channels (configurable with U7001 and U7005 input modules)
- synchronisation of two analysers allows up to 16 simultaneous channels/phases to be measured
- DC frequency band; 0.1 Hz to 5 MHz (15 MHz sampling)
- optional accessories for motor analysis, CAN/CAN FD interface and measured waveform output
- current (i.e. also power) measurement is only possible with dedicated HIOKI probes, e.g. 9272-05, CT684xA, CT686x-05, CT687x, CT687xA, CT6904A, etc.
- for more information about current probes, please contact us
- harmonic analysis up to 500th order
- unique automatic phase shift correction for highest measurement accuracy
- suitable for SiC and GaN semiconductor applications
- communication interfaces: LAN, GPIB, EXT I/O, optional CAN / CAN FD
- USB 3.0 for data storage
PW8001 analyzer - basic features and application examples: |
Recording of the webinar "Current trends and challenges in electrical power measurement": |
Downloads and links
Need some advice?
Call us on +420 325 610 123 and ask for a specialist in this field.